Agilent Technologies InfiniiVision 6000 Series User Manual page 168

Agilent technologies oscilloscopes user's guide
Hide thumbs Also See for InfiniiVision 6000 Series:
Table of Contents

Advertisement

4
Triggering the Oscilloscope
168
Frame (Start: Addr7: Read: Ack: Data: Ack: Data2) or Frame
(Start: Addr7: Write: Ack: Data: Ack: Data2) — The oscilloscope
triggers on a read or write frame in 7-bit addressing mode
on the 26th clock edge if all bits in the pattern match. For
triggering purposes, a restart is treated as a start
condition.
Read
SDA
SCL
Address
R/W
Start
Condition
10-bit Write — The oscilloscope triggers on a 10-bit write
frame on the 26th clock edge if all bits in the pattern
match. The frame is in the format:
Frame (Start: Address byte 1: Write: Address byte 2: Ack:
Data)
For triggering purposes, a restart is treated as a start
condition.
SDA
SCL
Address
Start or
1st byte
Restart
Condition
7 If you have set the oscilloscope to trigger on an EEPROM
Data Read condition:
Write
Ack
Data
Write
Address
Ack1
R/
2nd byte
Ack
Data 2
Trigger point
26th clock edge
Ack2
Data
Trigger point
26th clock edge
6000 Series Oscilloscope User's Guide
Ack
Stop
Condition
Ack
Stop
Condition

Advertisement

Table of Contents
loading

Table of Contents