Agilent Technologies InfiniiVision 6000 Series User Manual page 250

Agilent technologies oscilloscopes user's guide
Hide thumbs Also See for InfiniiVision 6000 Series:
Table of Contents

Advertisement

5
Making Measurements
250
and ∫ dt when 1-2 or 1+2 is the selected source if channel 1 and
channel 2 are set to dissimilar units in the channel Probe Units
softkey.
Math Measurements and Units
Only Peak-Peak, Maximum, Minimum, Average, X at Min, and X at Max automatic
measurements may be made on a math FFT function. See "Making time
measurements automatically" for the FFT X at Max and X at Min measurement.
Use the cursors to make other measurements on FFT. All voltage measurements
may be made on other math functions. The resultant units are:
FFT:
dB* (decibels)
2
2
1 * 2:
V
, A
or W (Volt-Amp)
1 – 2:
V (Volts) or A (Amps)
d/dt:
V/s or A/s (V/second A/second)
∫ dt:
Vs or As (V-seconds or A-seconds)
* When the FFT source is channel 1, 2, 3, or 4, FFT units will be displayed in dBV
when channel units is set to Volts and channel impedance is set to 1 M . FFT
units will be displayed in dBm when channel units is set to Volts and channel
impedance is set to 50 . FFT units will be displayed as dB for all other FFT
sources or when a source channel's units has been set to Amps.
The following figure shows the voltage measurement points.
Maximum
Digital channel voltage measurements
Automatic voltage measurements are not valid on digital channels on the
mixed-signal oscilloscope.
Amplitude
Peak-Peak
6000 Series Oscilloscope User's Guide
Top
Base
Minimum