Agilent Technologies InfiniiVision 6000 Series User Manual page 211

Agilent technologies oscilloscopes user's guide
Hide thumbs Also See for InfiniiVision 6000 Series:
Table of Contents

Advertisement

6000 Series Oscilloscope User's Guide
Figure 24
Cursors set to center of signal
7 Calculate the phase difference using the formula below.
second Y
--------------------------
sin
=
first Y
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is
the X-axis input, channel 2 is the Y-axis input, and channel 4 (or the External
trigger on 2-channel models) is the Z-axis input. If you only want to see portions
of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off
(analog oscilloscopes called this Z-axis blanking because it turned the beam on
and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V),
the trace is turned off.
Making Measurements
1.031
------------ - ;
= 37.65 degrees of phase shift
=
1.688
5
211