Agilent Technologies 5DX 5000 Series Installation Manual page 58

Table of Contents

Advertisement

Table B-1
Agilent 5DX Power Source and Loading Configuration Options
Power
Input Configuration,
Source
Input Wiring, and
and
Terminal Block Wiring
Loading
3-Phase
Delta
Load:
L1-L2
L1-L3
L2-L3
(17 A)
Input and Terminal Block
Wiring-Figure B-4
3-Phase
Wye Load:
L1-L2
L1-L3
L2-L3
(17 A)
Input and Terminal Block
Wiring-Figure B-4
3-Phase
Wye with
Neutral
Load:
L1-N
L2-N
L3-N
(10 A)
Input and Terminal Block
Wiring-Figure B-6
B-50 Wiring Configuration
L1
L2
L3
G
L1
L2
L3
G
L1
G
N
L2
L3
Description
200 - 240V
Three Phase
208 - 220V
Three Phase
Wye
380- 415V
Three Phase
Wye w/Neutral
VAC
200V
220V
230V
240V
120/208V
127/220V
220/380V
230/400V
240/415V
Agilent 5DX Installation