X-ray Start-up
7.
similar to Figure 16. Press Esc to exit X-ray Startup, since you don't
want to bring up X-rays yet.Panel Handling should be verified before
X-rays are turned on.
COLD Start
Figure 16
X-ray Start-up Screen
NOTE
Other error or warning messages could occur during this process.
In general, press Esc to clear these. If error dialog boxes appear
with a choice, use one of the choices. For example: "Failed to
establish X-ray power on, do you wish to retry start-up (Y/
N)". Press N to disable the retry.
Agilent 5DX Installation
- The system should now bring up a start-up screen
5DX X-ray Control
Agilent Technologies
Kilovolts (160)
0
Focus (V)
Grid (V)
765.0
500.0
Dialog
X-RAYS were down for an extended period (cold start)
To continue, turn X-ray Control switches on and
press <Enter>. To quit, Press <Esc>.
xx minutes remaining
MicroAmps (100)
0
Filament (V)
0.0
29