POWER
Figure 12
Vacion Power Supply
Check Interlocks
4.
Tap each of the 6 side doors, 2 topside access panels, tower door and
interlock points with your hand. Watch to see if the red interlock LED
on the control panel goes off. If it does, adjustment of interlock
switches may be required. (Refer to the Service Manual, Safety
Interlock Subsystem section for more information.).
Figure 13
Operator Control Panel
26
VACION PUMP
POWER SUPPLY
INPUT
~200-240 V 60VA
ON
— Enable the X-ray key and press the
Measure the voltage
here (J04000)
J04000
Watch the interlock LEDs to
make sure they stay on.
switch.
ON
Agilent 5DX Installation