Move 5DX To Install Site
5.
the four (15/16") nuts from the 5DX in the bottom four corners of the main
frame. When you are ready, (using acceptable lifting equipment) raise the
5DX from its pallet and move to the install area. You will need assistance to
move the tethered electronics bay along with the main unit.
Caution
Be very careful while moving the forklift in place to remove the
5DX from its pallet. You will likely have to do this at the rear of
the unit and there are wires, air lines and the interlock controller
that can all be damaged by careless forklift use.
Staging
6.
beginning installation, find the box with the foot pads and spacers
(part number E7200-80024).
Clear Packaging
7.
pallet. Remove the remaining packing materials, straps, padding etc.
from the 5DX. Tape is applied to the edges of the tower door to
prevent damage during shipping. Be sure to remove this tape to ensure
a good door seal.
Inspect the 5DX
8.
Inspect system exterior for abrasions or dents. Record any damage on
the waybill. Advise the carrier and contact
Services
WARNING
While lowering the 5DX on the threaded foot pads and spacers,
make sure the forklift operator is aware you will be under the
raised system.
System Placement
9.
install area, move it to the desired placement location. Position the
system according to the customer's requirements then install the foot
8
— Organize all materials in the installation area. Prior to
— Remove all remaining items from the 5DX
— Inspect the 5DX and accessories for damage.
immediately.
— When the system has been delivered to the
To remove the 5DX from its pallet, remove