Load the confirmation and adjustment board, if not still in from
b.
previous step.
From the
c.
choose
A Caution Box will appear:
d.
Board will need to be exposed to continuous X-rays. Do you want to
turn on Live X-rays (Y/N)
Press Y.
Follow the steps on the screen: verify that the alignment points are in
e.
range. Perform adjustment if necessary, including the Z-axis for
focus. Be aware that large corrections in any of the axes indicate a
possible change in an axes' zero reference, or change in rail location.
If the alignment is not where expected, it may mean that an
NOTE
alignment or reference in the axes assemblies has changed. If
alignment is significantly moved (such as out of the view), verify
the confirmation & adjustment panel is loaded correctly. Home
the axes and recheck. Next, try a customer board and verify that
the alignment is off by the same amount. If the alignment is
significantly out of range, it may require a verification of the axes
assemblies' alignment. Refer to the System Administration
Manual.
Select Automatic. Verify that the images are clear and focused.
f.
From the
g.
From the top level
h.
Provide data entries as requested (for most items, simply pressing
Enter will be appropriate).
Review the measurement information and verify there were no
i.
failures.
Agilent 5DX Installation
Utilities
menu, select
Manual
Use Previous Alignment for all views
and
Utilities
menu, select
User Interface
Align
Board. When prompted,
Surface
Map. Select Automatic.
menu, select Test. Select Panel.
37