the labels. Be very careful to when making this connection since the
pins are very small and fragile. Use great care to orient the sockets
correctly! Incorrect connection CAN BE forced. Connector
damage will result. If the sockets are incorrectly oriented, there will
be no image in the image window. If Data1 and Data2 are switched,
the picture on the image window will be upside down.
Connect the
c.
Connect the shutter cable (
d.
the
SYNC
closest to the LEDs).
AC Main Hookup
17.
main power installation to local code. For general service feed
location, see Figure 8 on page 17 and Figure 10 on page 22. Normal
requirements are:
• Between 195 and 240 volts, three phase (3Φ) 5Ø or 6Ø Hz. (See
appendix C for overall requirements)
• Where allowed, the recommended attachment is via flexible conduit,
allowing a range of motion for the electronics bay of a minimum of 1
meter (a bit over 3 feet) to allow for serviceability. For optimum
flexibility, extend this to 2 meters. Cables must be the stranded type
(not solid). Follow local code if it supersedes these instructions.
• A breaker panel near the 5DX installation area can be used to allow
complete power down in case of an emergency. The E1135C PDU
(Power Distribution Unit) has a rotating, lockable switch for this as
well.
5DX systems are pre-wired with a cord and plug. The local
NOTE
electrician will only need to install a suitable outlet where the
plug can be inserted. Follow local code regulations.
Agilent 5DX Installation
(telephone style) connection on the camera.
SERIAL
CAMERA SHUTTER, E7200-61463
connector on the camera (the sync connector is the one
— Work with the customer electrician to provide
) to
21