problem--do not bring up X-Rays. Refer to the online Service Guide
(N7200-90041) for troubleshooting.
Caution
If the reading is not less than 20mv, allow more time for the
Vacion power supply to operate. Recheck this again before
proceeding.
Select
b.
From the DOS prompt, type
c.
Type
d.
have a value of
Caution
Failure to do this step could result in permanent damage to
the X-ray tube. The X-ray tube has components that need to
warm up before generating X-rays. The system has been un-
powered for an extended period of time so the X-ray tube requires
a 60 minute start up.
After editing the xray.cfg file, use Alt W to save the file and Alt X to
e.
exit the editor.
From the Operator Control Panel (Figure 13 on page 26) turn the X-
f.
Ray Enable Key to
From the DOS prompt, type
g.
directory. Then type
warning will be displayed. Press Esc to continue. The time
remaining should indicate 60 minutes in the status display.
After 60 minutes ramp time, the high voltage should show about 160
h.
kv in the status display and remain stable. If it does not, there is a
problem and you should contact Agilent 5DX Hardware Support.
34
System Access
vs xray.cfg
instead of
60
Enable
xrayon -zero
then
Quit from
and press Enter.
ca
and edit the line
Cold Start Rampup
.
15
, then push the X-Ray
and press Enter to enter the master
ma
5DX.
Button.
On
and press Enter. A
Agilent 5DX Installation
to