Select the camera icon on the graphic display window. Switch the
g.
display to Live Laminography.
Verify that the 5 dot calibration pattern is visible on the graphic
h.
display window. (Some adjustment of the focus (Z axis) may be
required.)
From the DOS prompt type
i.
Self Test Validation
9.
properly, perform the following steps:
From the
a.
Select
b.
Select
c.
Test Setup
Go to
d.
menu, select Run.
For information on running self tests, see the System Administration
e.
Guide (on line with the system manuals).
Verify that all tests pass. If tests do not pass, refer to the Service
f.
Guide (N7200-90041) or try self test setup and run self test again.
It may be difficult for Self-Test to pass until Confirm and Adjust
NOTE
has been run.
System Operation Functional Checks
10.
system is functioning properly, perform the following steps:
From the
a.
choose
36
— To verify that Self Test is functioning
System Access
Automatic Startup
Utilities > Service Menu > Adjust > Self-test > Self
Utilities > Service Menu > Adjust > Self-test.
Utilities
menu, select
. Select a 400 FOV.
5DX_512
and press Enter.
auto
menu select
System
to bring the 5DX to a ready state.
— To verify that the overall
Load Panel Program
Startup.
From the
then
Agilent 5DX Installation