Thermo Scientific Partisol 2000i Instruction Manual page 73

Air sampler
Table of Contents

Advertisement

Converting a Partisol
2000i from a PM
Sampler to a PM
Sampler
Thermo Fisher Scientific
Figure 4–12. PM
The Partisol 2000i Sampler is now configured for PM
no difference in Sampler programming or operation with the Sampler
configured as a PM
and calculates sample volumes in Standard and Volumentric conditions
regardless of the Sampler configuration.
Note Make sure that the filter cassettes, tranport containers, and data
sheets are clearly labeled as PM
you install the PM
Also ensure that the proper filters are used for the type of sampling being
performed.
PM
is reported to a standard temperature and standard pressure. Set the
10
standard temperature to values appropriate for the specific location and
agency. The U.S. EPA requires that these values be 25 °C for temperature
and the standard pressure be 760 mmHg. Refer to
on page
9-24
and
Converting a PM
sampling can be accomplished by removing the PM
10
tube, and installing the VSCC, or WINS impactor, and adapter. Follow
the Steps for converting a Sampler from a PM
2.5
configuration, but in reverse order as described below.
Use the following procedure to convert a PM
Sampler.
Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 Sampler
Inlet on the Sample Tube
10
or PM
Sampler. The Sampler properly measures
10
2.5
sampling filters or sampling data when
10
inlet and straight tube on the Partisol 2000i Sampler.
10
"Standard
Pressure" on page 9-25.
configured Sampler to a Sampler configured for PM
10
Partisol 2000i and Partisol 2000i-D Instruction Manual
Partisol 2000i Inlet Conversion
sampling. There is
10
"Standard
Temperature"
pass through adapter
10
configuration to a PM
2.5
Sampler to a PM
10
2.5
2.5
10
4-7