AK Protocol
Commands List
E-10
Partisol 2000i and Partisol 2000i-D Instruction Manual
RS-Para 1
52
ASCII code for the 1-digit Station Number (for example "4": 052). The Station Number is always 1 digit in length.
ASCII code representation of the 2-digit Channel Number (for example: "K0": 075, 048). The Channel Number is
RS-Para 2
75048
always 2 digits in length.
Optional: Up to 3 ASCII codes can be added to response from instrument. In this case, <CR> and <LF> (ASCII
RS-Para 3
13010
codes 013 and 010) are appended to the response. Enter 0 if nothing is to be appended.
RS-Para 4
0
Not used.
Transmission to Instrument
Byte
Example
1
<STX>
ASCII code 002.
2
4
1-digit Station Number, RS-Para 1.
3
S
4
F
Set Function command, where xx
represents a 2-digit code. These
codes are defined below.
5
x
6
x
7
<space>
Space.
8
K
2-digit Channel Number, as
defined by RS-Para 2.
9
0
10
<ETX>
ASCII code 003.
LISTING OF FUNCTION CODES (xx)
01
Run
02
Stop
10
Set Time
11
Set Date
50
Switch to "None" RS232 Mode
AK Protocol
Set Function Command (SFxx)
COM 2-WAY SETTINGS
Description
B
No Err
1
<STX>
2
3
4
5
6
7
<space>
8
9
<ETX>
10
<CR>
11
12
13
14
15
16
17
18
19
Response from Instrument
Error
<STX>
ASCII code 002.
4
4
1-digit Station Number, RS-Para 1.
S
S
F
F
4-digit Set Function command, with
the 2-digit xx code corresponding to
the function that was set.
x
x
x
x
<space>
Space.
0
0
Number of current status conditions.
<space>
Space.
S
Up to 3 digits appended to the end
<LF>
E
of the response transmission,
according to the entry for RS-Para 3.
<ETX>
<CR>
<LF>
Description