Thermo Scientific Partisol 2000i Instruction Manual page 311

Air sampler
Table of Contents

Advertisement

Thermo Fisher Scientific
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig in
Receive:
dig in 0xff7f
din channel
This command reports the action assigned to the digital input channel and
the index number of the corresponding active state. The following example
reports the input 1 to be assigned an index number 1 corresponding to
action of STOP with the active state being high.
Send:
din 1
Receive:
din 1 1 STOP high
set din channel index state
This command assigns digital input channel (1-16) to activate the action
indicated by index (1-12), when the input transitions to the designated state
(high or low). Use the "list var din" command to obtain the list of
supported index values and corresponding actions. The following example
sets the digital input channel 5 to 1 on a low-to-high transition.
Send:
set din 5 1 high
Receive:
set din 5 1 high ok
dout channel
This command reports the index number, output variable and the active
state assigned to digital output channel. The following example reports
output 4 to be assigned an index number 4 corresponding to action of "I/O
board status".
Send:
dout 4
Receive:
dout 4 4 I/O BD STATUS open
set dout channel index state
This command assigns digital output channel (1-10) to be assigned to the
action associated with index (1-34), and assigns it an active state of state
(open or closed). Use the "list var dout" command to obtain the list of
supported index values and corresponding state. The following example sets
the digital output channel 4 to state 11.
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
Partisol 2000i and Partisol 2000i-D Instruction Manual
C-Link Protocol Commands
I/O Configuration
B-31

Advertisement

Table of Contents
loading

This manual is also suitable for:

Partisol 2000i-d