Thermo Scientific Partisol 2000i Instruction Manual page 288

Air sampler
Table of Contents

Advertisement

C-Link Protocol Commands
Measurements
B-8
Partisol 2000i and Partisol 2000i-D Instruction Manual
ambient rh
This command reports the current value for ambient relative humidity, in
percent.
Send:
ambient rh
Receive:
ambient rh 75.5
ambient temp
This command reports the current value for the ambient temperature in
degrees C.
Send:
ambient temp
Receive:
ambient temp 24.1
ecomp temp
This command reports the value for the electronics compartment
temperature in degrees C.
Send:
ecomp temp
Receive:
ecomp temp 30.3
fcomp temp
This command reports the filter compartment temperature in degrees C.
Send:
fcomp temp
Receive:
fcomp temp 20.4
pump pres
This command reports the pump pressure in mmHg.
Send:
pump pres
Receive:
pump pres 800.0
filter pres
This command reports the value for the filter pressure in mmHg.
Send:
filter pres
Receive:
filter pres 765.4
filter2 pres
This command reports the value for the filter2 pressure in mmHg.
Send:
filter2 pres
Receive:
filter2 pres 734.5
filter temp
This command reports the value for the filter temperature in degrees C.
Send:
filter temp
Receive:
filter temp 18.8
Thermo Fisher Scientific