Typical Configuration Example - HBM DMP41 Operating Manual

Digital precision measuring
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

Setting mode
DMP41
7.6

Typical configuration example

F1
The transducer is a load cell with the following nominal (rated) data:
maximum capacity: 50 kg, sensitivity: 2.5 mV/V
The weight of a small amount of bulk material in a container is to be
determined.
► Connect the load cell to the "INPUT B" socket
"Back of device", Page 22.
Channel configuration
► Use shift key F1 to select setting mode.
► Under Range (bridge excitation voltage), choose the selection fields 5 V
and 2.5 mV/V (A).
► Press Save(diskette symbol).
► Press more....
► Press Scaling and select the unit kg (C).
► Use the keypad to make entries in the edit fields (D):
Point 1: 0.000 kg and 0.000 mV/V
Point 2: 50.000 kg and 2.500 mV/V
► Press Save (diskette symbol).
Range
A03470_07_E00_00 HBM: public
Typical configuration example
è
see Chapter 3.4.2
A
Scaling
D
Configuration
C
97