Error messages/remedial action
102
Filter settled
The channel low-pass filter is settled, it deviates less than 1 ppm from the
full scale value. Actions such as zeroing or taring are only possible in a
settled state.
The settling process is still ongoing.
Sense line 2/3 detected (Sense line 2/3)
The sense line is present and properly connected, no errors are present.
The sense line connection has not been detected. Either no transducer is
connected or the sense line connection is missing
Signal in the measuring range (Signal within range)
The measurement signal is within the measuring range, there is no over­
loading.
The amplifier is overloaded, the measured value is invalid. Overflow is
displayed in the channel display.
Excitation
The excitation voltage is within the required range, no errors are present.
The excitation voltage has been switched off, e.g. because the transducer
resistance is too low or a short-circuit is present.
TID
A TEDS module has been detected, the module ID has been read out. The
TEDS information is not yet evaluated in the current version, only the
(worldwide) unique ID is read out.
No TEDS module has been detected.
A03470_07_E00_00 HBM: public
.