Configuration
Configuring measurement channels in setting mode
66
7.2
Configuring measurement channels in setting
mode
F1
Important
The measurement channels are set via the menus under Channel.
If the device is in measuring mode:
► Press shift key F1 to switch to setting mode (Settings).
7.2.1
Measuring range (Range) and bridge excitation voltage
(Excitation)
The measuring range selection is dependent on the choice of bridge
excitation voltage.
Bridge excitation voltage (V)
2.5
5 (factory setting)
10
► Press Range.
► In the Range and Excitation menu, choose the required bridge
excitation voltage/measuring range combination.
► Press Save.
Important
The maximum accuracy as stated in the specifications is only achieved with
settings of 10 V for the excitation voltage and 2.5 mV/V for the measuring
range.
The bridge excitation voltage selection can restrict the choice of measuring
range.
A03470_07_E00_00 HBM: public
Selectable measuring range (mV/V)
5
10
2.5
5
2.5 (factory setting)
DMP41