Configuration
Configuring measurement channels in setting mode
72
7.2.5
Measured
peak value
The peak values function (Min/Max) can be used to record and save iso
lated signal peaks and minimum/maximum signal amplitudes. All peak val
ues are stored continuously.
You can display min/max and peak-to-peak as gross, net, or absolute val
ues, either in mV/V or scaled as defined by the user, or query them using
the MSV command.
Gross
With fast dynamic signals, you must take into account that the peak values
are determined from the filtered signal.
If, in the above example, the peaks of the the foundation oscillations are to
be measured, the filter must be set to more than 12 Hz.
A03470_07_E00_00 HBM: public
Peak-value memory (Min/Max)
Peak value memory
Display
DMP41