Safety Features - Otto Bock 7000P Instructions For Use Manual

Litelift pro cpm device
Table of Contents

Advertisement

Danger!
Before raising the CPM upright for storage, place the CPM into full flexion to avoid foot plate interference with
mounting bar or bed frame.
Danger!
When applying the CPM to patient, make sure that the CPM is fully supported on the bed: the CPM should
NOT be supported by lift mechanism while applied to patient.
Danger!
The controller must be positioned so that patients and medical personnel can reach the START/STOP button
during use.
Attention!
Use the device only in accordance with the Physician prescription and these Instructions for Use. Failure to do
so may result in damage to the device and/or personal injury.
NOTE: Patient Kits are for single patient use only.
Attention!
The use of accessories and cables other than those specified, with the exception of accessories and cables
qualified and sold by the manufacturer of the equipment, may result in increased emissions or decreased
immunity of the equipment and may cause the system to be non-compliant with the requirements of IEC 60601-
1-2:2004.
Attention!
The equipment should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, verify operation in that configuration.
Attention!
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
Attention!
Use only manufacturer's supplied replacement components.
Attention!
Do not use the device if there are mental or physical conditions that preclude patient compliance.
Attention!
Do not expose the device to water or extreme temperatures.
Attention!
Do not use the device near exposed flames, while smoking or near excessive heat.
Turn the power off before unplugging.
Attention!
Select a location for the device and device components (controller, straps, cables and power supply, if
applicable), to prevent a tripping hazard during use.
Attention!
Device must be stored in a location that has a least 19 inches of clearance at all times.

2.3 Safety Features

Low Voltage
The power supply delivers less than 20 volts DC to the device. The LiteLift PRO will tolerate electrical supply variations
which may be found in the home or hospital environments.
4