6.2.4 Joint Set-up Screen: Setting CPM Device Parameters
Force, Pause, Pause Time and Speed Parameters for the specific joints in a CPM device are set in the Joint Set-up
Screen (Fig. 44). To set these parameters, use the Soft Keys to navigate to the desired Joint Set-up Screen if there is
more than one joint involved.
Fig. 44
Knee Joint Setup
Selection Cursor
selected
Fig. 45
Knee Joint Setup
Slider Bar selected
Next, use the Direction Keys to move the Cursor to the desired parameter (Fig. 44), and press the Select Key to select
that parameter. A Setting Slider Bar will appear (Fig. 45) that will allow you to change the setting with the left and right
Direction Keys. Once you have chosen your setting, press the Select Key to set, and scroll to the next parameter setting.
Range of Motion (ROM) parameters Use the left and right Direction Keys to navigate to the ROM Setting Column
(Fig. 43, Pos. E, page 24). When the PRO Controller is first connected to a CPM device it will display the CPM device's
ROM limits in the Setting Limit windows (Fig. 43, Pos. C, G, page 24). Both these and the Patient Controlled ROMs
(Fig. 43, Pos. D, F, page 24) can be set using the keypad. Simply scroll to the desired setting and select, then use the up
and down Direction Keys to raise or lower the slider bar to the desired degree.
Flexion
Force
Pause
Speed
Force
Pause
Speed
Extension
Flexion
Force
Pause
Speed
Force
Pause
Speed
Extension