During Use - Zeiss KSK 150 FC Instructions For Use Manual

Table of Contents

Advertisement

KSK 150 FC

During use

Version 7.1
G-30-1158-en
• Make sure the operating staff has been briefed on the use of the device.
• Never cover any ventilation openings. This may cause the light source of
R077
the device to overheat and fail.
• Make sure that all symbols and labels on the instrument are legible and
understood by operating personnel.
• Carry out the functional test described in the section "Operation" before
every use of the device.
• Only use the system after it has been properly balanced.
• Never operate the system unattended.
• Avoid looking directly into the light source, e.g. into the objective lens or
light guide.
• Start setting the lamp brightness at the lowest value and gradually in-
crease the brightness until the image is sufficiently bright.
• Defective accessories or accessories not approved by Carl Zeiss may lead
to increased leakage current!
Never touch signal interfaces while in contact with the patient.
• The main plug must only be connected to a socket with a faultless protec-
tive ground conductor.
• When setting up your system, make sure to comply with the requirements
of IEC 60601-1-1:2000 and IEC 60601-1:2005 + A1:2012, Chapter 16.
• Do not pull at the light guide, power cord or any other connecting cables.
• If you need to change the lamp shortly after it has failed, the lamp will still
be very hot. Use heat-resistant protective gloves when replacing the lamp
to prevent burns.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions", attach a sign to the device stat-
ing it is out of order and contact our service representative.
• Please note that the interaction of heat and antimicrobial substances in in-
cision foils may lead to an increase in the patient's reaction to these sub-
stances.
Safety Measures
Page 13