KSK 150 FC

Disinfection

CAUTION
R070
Version 7.1
G-30-1158-en
• Please observe the Instructions for Use supplied with each anti-fogging
agent.
Anti-fogging agents not only ensure fog-free eyepiece optics, they also clean
the eyepiece optics and protect them from dirt, grease, dust, lint and finger-
prints.
Cleaning mechanical surfaces
All mechanical surfaces of the system can be cleaned by wiping them with a
damp cloth. Do not use any aggressive or abrasive cleaning agents.
Clean off any residue using a mixture of 50% ethyl alcohol and 50% distilled
water plus a dash of household dish-washing liquid.
It may be a requirement to disinfect surfaces.
Surfaces damages on device!
• Use a disinfectant based on aldehyde and/or alcohol. The addition of
quaternary compounds is acceptable. To prevent damaging surfaces, dis-
infecting components other than those listed below must not be used.
The maximum concentrations are:
For alcohol (tested with 2-propanol): 60%
For aldehyde (tested with glutaraldehyde): 2%
For quaternary compounds (tested with DDAC): 0.2%