KSK 150 FC
Colposcope
Procedure
CAUTION
R035
Version 7.1
G-30-1158-en
Check that the binocular tube is screwed on tightly to the colposcope
body with the securing screw.
Check that the objective lens is screwed on tightly.
• Turn on the device with the power switch of the suspension system.
Eyepieces / binocular tube
Check that the colposcope and the tube are in an ergonomic position.
Check that the correct interpupillary distance has been set.
Check that the eyecups have been adjusted in such a way that you can see
the full field of view.
Check that the correct prescription has been set on the diopter scale.
Check that image quality is correct throughout the entire magnification
range.
Check that the fine focusing can be adjusted.
Accessories
Proper functioning of further device components has been checked using
the relevant Instructions for Use.
Risk of eye injury!
Looking directly into the light source, e.g. into the objective lens, may result
in damage to the eye.
• Avoid looking directly into the light source!
• Do not use this device for ophthalmic procedures.
Operation
Page 53