Changing Microscope Accessories; Setting The Balance - Zeiss KSK 150 FC Instructions For Use Manual

Table of Contents

Advertisement

Preparations for use

Changing microscope accessories

Setting the balance

NOTE
R049
R076
Page 44
The colposcope accessories can be changed in reverse order to that described
in "Attaching binocular tube, eyepieces and objective lens". Please observe
the following:
• Turn off the unit at the power switch before changing any accessories.
• Bring the suspension arm in a position convenient for you.
• Firmly tighten "Brake" adjusting knob (1) to lock the arm in position.
• After changing the accessories, re-adjust the friction as required using the
"Brake" adjusting knob (1).
Rebalance the system after every change of equipment.
Risk of injury caused by lowering of the colposcope!
An incorrectly balanced colposcope may move downwards unexpectedly and
cause injuries.
• Observe the maximum load capacity of the suspension arm. See the ap-
propriate warning label on the instrument head.
• Use the colposcope in properly balanced condition only.
• Use approved accessories only.
• Position the colposcope in such a way that you can move it away from
the patient at any time.
• Hold the suspension arm with your hand and release the brake using the
"Brake" adjusting knob (1).
• To set the balance, turn the "Balance setting" adjustment knob (2) until
the suspension arm, with its brake released, can be smoothly moved up-
wards and downwards in the entire adjustment range and stops in any re-
quired position.
KSK 150 FC
Version 7.1
G-30-1158-en

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents