Operating Instruction STERIclave B 2200 / 2200 P
|
7 Maintenance
7.4 Regular total cleaning
+
CAUTION
▶ Carefully screw the filter into place by hand.

7.4.3 Clean the chamber

Requirement
The unit is in "OFF" mode.
Note
Only KaVo cleaning tabs (Mat. no. 1.004.8678) may be used.
▶ Place a detergent tab inside the chamber and close the doors.
▶ Hold down the Start/Stop button, and then pressure the Power button.
Automatic cleaning starts.
"CLEANING" will appear on the display.
The vacuum pump is running.
Vapour may escape when you open the door.
Risk of scalding
▶ Do not open the door during automatic cleaning.
Automatic cleaning takes approx. 15 minutes and cannot be interrupted.
The Ready LED will light up as soon as automatic cleaning is complete.

7.4.4 Changing the water

See also: 6.10 Changing the water, Page 54
62/78