Operating Instruction STERIclave B 2200 / 2200 P
|
6 Operation

6.6 Tracking the program sequence

Phase
Display/LED/
Acoustic signal
Program display
Approx. 10 seconds
Vacuum phase
Vacuum lights up
5-10 minutes
Sterilisation phase
Sterilize lights up
Decompression
Sterilize lights up
End of sterilisation
Sterilize flashes
Aspiration of residual
Drying lights up
vapour
Drying
Drying lights up
Airing
Drying lights up
End of drying phase
Ready and Sterilize
(class B sterilisati‐
light up + acoustic
ons)
signal sounds
End of drying phase
Ready lights up +
(Fast cycle and Spe‐
acoustic signal
cial S4)
sounds
6.6 Tracking the program sequence
Time display
Duration of selected
program
Remaining time
in phase
Remaining time
in phase
Remaining time
in phase
Remaining time
in phase
Process duration
Remaining time
in phase
Remaining time
in phase
Total duration of pro‐
cess
Total duration of pro‐
cess
If the door is opened within half an hour of the drying phase finishing, the heating
will remain in the pre-heating phase.
Pressure display
Pressure of selected
program
Current pressure
Current pressure
Current pressure
(falls back to zero)
Current pressure
Current pressure
Current pressure
Current pressure
Current pressure
Current pressure
50/78
Temperature display
Temperature of se‐
lected program
Current temperature
Current temperature
Current temperature
(drops)
Current temperature
Current temperature
Current temperature
Current temperature
Current temperature
Current temperature