Non-Specific Error Messages - KaVo STERIclave B 2200 Operating Instruction

Table of Contents

Advertisement

Operating Instruction STERIclave B 2200 / 2200 P
|
8 Troubleshooting

8.5 Non-specific error messages

Malfunction
OPEN DOOR
DRY FAIL
ADD H2O
FULL H2O
MANU STOP
BLAC OUT
NEED CLEANING
NEED SERVICE
NEED INST
TEST FAIL
The doors do not open
8.5 Non-specific error messages
Cause
The door was not opened at the end
of the program sequence.
Started with door not closed proper‐
ly or open.
Items removed prior to the end of
the drying program. The items have
been sterilised but have not dried
completely.
No more water in fresh water tank.
(This message will appear prior to
the start of the program.)
The used water tank is full. (This
message will appear prior to the
start of the program.)
The program has been interrupted
by means of manual intervention.
Sterilisation has not been comple‐
ted.
Blackout during process.
60 cycles without intermediate clea‐
ning.
Maintenance must be carried out af‐
ter a year or 2000 cycles.
Vacuum test failed.
"OFF" is not shown on the display. ▶ Press the Start/Stop button.
"OFF" is shown on the display.
The locking lever is horizontal.
The lever cannot be moved.
Sterilisation is over, and the green
"Ready" LED shines.
Remedy
▶ Open the door.
▶ Close the door properly.
▶ Press the Start/Stop button
▶ Add more distilled water.
▶ Empty tank (maintenance may
▶ If there is moisture inside the
▶ Check the mains plug.
▶ If there is moisture inside the
▶ Run a cleaning cycle.
▶ Have the unit serviced by a qua‐
The message will disappear at the
start of a cycle and reappear the
next time the unit is switched on.
▶ Run installation.
▶ Clean the door seal and repeat
▶ Contact customer service.
▶ Turn on the unit, Press the
▶ Pull firmly on the doors.
▶ Turn the lever all the way to the
▶ Press the Start/Stop button, and
69/78
be required).
chamber, dry the chamber and
repeat the program.
chamber, dry the chamber and
repeat the program.
lified engineer.
the test.
Power button, and wait until the
self-test is over.
right, and then turn it to a hori‐
zontal position.
then move the lever into a hori‐
zontal position.