Dansensor
LeakPointer H
®
Replacing the gas springs
NOTE! We recommend that two persons perform this operation, as the lid must be
held open while the gas springs are removed and mounted, respectively.
Required tools:
Flat tip screwdriver, small
1. Drain the tub
2. Disconnect the drain hose
3. Remove the original gas springs
locks
in both ends and press them downwards (upwards at the top) until you are able
to pull the springs off the pivots
4. Mount the new gas springs (before that the pivots
them as described above).
5. Before mounting the springs, return the locks
gas springs onto the pivots.
O
EN
2
as described in
"Cleaning procedure" on page
then remove the tub
. To do so, use a flat-tip screwdriver
(see detail).
22.
.
must have been removed from
to their original positions, then click the
User Guide
27
to release the
P/N 390225-C
11/2021
Need help?
Do you have a question about the mocon Dansensor LeakPointer H2O and is the answer not in the manual?