4
General Maintenance
Pull on the connectors, not on the wires.
C A U T I O N
4 Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Disconnect them there.
5 Remove the thumbscrews that hold the ion source in place.
6 Pull the ion source out of the source radiator.
The analyzer operates at high temperatures. Do not touch any part until you are sure
WA R N I N G
it is cool.
5975T LTM GC/MSD Operation Manual
133