wires which can conduct ESD to sensitive components on the side board. ESD
damage may not cause immediate failure but will gradually degrade
performance and stability. See
Electrostatic discharges to analyzer components are conducted to the side board
C A U T I O N
where they can damage sensitive components. Wear a grounded antistatic wrist strap
(see
chamber.
Some analyzer parts should not be disturbed
The mass filter (quadrupole) requires no periodic maintenance. In general, the
mass filter should never be disturbed. In the event of extreme contamination,
it can be cleaned, but such cleaning should only be done by a trained Agilent
Technologies service representative. The HED ceramic insulator must never be
touched.
Incorrect handling or cleaning of the mass filter can damage it and have a serious,
C A U T I O N
negative effect on instrument performance. Do not touch the HED ceramic insulator.
More information is available
If you need more information about the locations or functions of analyzer
components, refer to the Agilent 5975T LTM GC/MSD Troubleshooting and
Maintenance Manual. Many procedures in this manual are illustrated with
video clips.
5975T LTM GC/MSD Operation Manual
page
115) and take other antistatic precautions before you open the analyzer
page 115
for more information.