4
General Maintenance
Septum retainer nut
Septum
Insert assembly
4 Firmly press the new septum or Merlin Microseal into the fitting. The metal
5 Install the septum retainer nut or Merlin cap and finger-tighten. Tighten the
6 Restore the analytical method.
7 Reset the septum counter.
122
parts side of the Merlin Microseal should face down (toward the oven).
septum retainer nut until the C-ring is about 1 mm above the nut.
1 mm