Operating in Electron Impact (EI) Mode
5 Click Apply to download setpoints or click OK to download setpoints and
6 To make the new settings part of the method, select Save from the Method
Make sure that the carrier gas is turned on and the column has been purged of air
C A U T I O N
before heating the GC/MSD interface or the LTM column.
82
The typical setpoint is 280 °C. The limits are 0 °C and 350 °C. A setpoint
below ambient temperature turns off the interface heater.
close the window.
menu.
5975T LTM GC/MSD Operation Manual