Operating in Electron Impact (EI) Mode

To Set Monitors for MSD Temperature and Vacuum Status

A monitor displays the current value of a single instrument parameter. They
can be added to the standard instrument control window. Monitors can be set
to change color if the actual parameter varies beyond a user-determined limit
from its setpoint.
Procedure
1 Select MS Monitors from the Instrument menu.
2 In the Edit MS Monitors box, under Type, select Zone.
3 Under Parameter, select MS Source and click Add.
4 Under Parameter, select MS Quad and click Add.
5 Under Parameter, select TurboSpd and click Add.
6 Select any other monitors you want and Add them.
7 Click OK. The new monitors will be stacked on top of each other in the lower
8 Click and drag each monitor to the desired position. See
Figure 18
9 To make the new settings part of the method, select Save from the Method
78
right corner of the Instrument Control window. They must be moved for you
to see them all.
way of arranging the monitors.
Arranging monitors
menu.
Figure 18
5975T LTM GC/MSD Operation Manual