Installing LTM Columns
7 Position the column so it extends 4 to 6 mm past the end of the ferrule
Figure 6
8 Slide the septum to place the nut and ferrule in the correct position.
9 Insert the column in the inlet, and finger-tighten the nut.
10 Tighten the column nut an additional 1/4 to 1/2 turn. The column should
11 Carefully wind the guard column clockwise around the three column
12 Check the Guard Column Enclosure to be sure that the column does not
36
(Figure
6).
Capillary column
Ferrule (inside nut)
Inlet column nut
Septum
Installing a capillary column for a split/splitless inlet
not slide with a gentle tug.
brackets. See
Figure
touch the heated walls.
7.
4 to 6 mm
5975T LTM GC/MSD Operation Manual