General Maintenance
Procedure
1 Turn off the inlet heater and flow.
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot,
WA R N I N G
wear heat-resistant gloves to protect your hands.
2 Slide the locking tab forward (counterclockwise). Lift the septum assembly
3 Loosen the O-ring from the sealing surface with tweezers.
124
straight up and away from the inlet to avoid chipping or breaking the liner.