Installing LTM Columns
Always wear safety glasses when handling capillary columns. Use care to avoid
WA R N I N G
puncturing your skin with the end of the column.
Wear clean, lint-free gloves to prevent contamination of the parts.
C A U T I O N
2 Slide a septum, column nut, and conditioned ferrule onto the free end of the
Capillary column
Column cutter
Ferrule, taper up
Inlet column nut
Septum
Figure 4
3 Use the sharp edge of the column cutter to score the column 2 cm from the
34
column
(Figure
4). The tapered end of the ferrule should point away from
the column nut.
Preparing a capillary column for installation
end. See
Figure
5.