To Remove the RIS Probe Extraction Tool from the Instrument

To Remove the RIS Probe Extraction Tool from the Instrument

WARN IN G
104
This procedure is part of an automated series of steps
controlled by MassHunter. Do not attempt this procedure unless
prompted by MassHunter. The RIS Status area on the GC/Q-TOF
Tune tab must show the Gate Valve is Open and the RIS vacuum as
0.000 Torr.
The 7200 Q-TOF GC/MS operates at high temperatures. Make sure
the RIS cooling chamber is at a safe temperature before attempting
to remove it from the instrument.
1
Undo the latch on the right side of the RIS probe extraction
tool and open the door.
2 Lift the RIS probe extraction tool up to release it from its
hinges.
3 Replace the RIS probe extraction tool with the RIS chamber
door.
4 Line up the hinge pins on the RIS chamber door with the
slots on the instrument hinge, and lower the pins into the
hinge.
5 Swing the RIS chamber door closed and secure the door with
the latch on the right side.
6 Return to
page 94.