■D/A output select

Hold, Realtime, Zero scale, Full scale
Hold
Realtime
Zero scale
Full scale
How to set
SET.
System Setting
How to output adjust
Select each fixed output in D/A output select, and press the
simulated trimmer appearing on the adjustment screen to adjust the output.
While monitoring the output value, make fine adjustment of the output with the UP/DOWN
button of the simulated trimmer, and determine by pressing the
that the trimmer position is entered.
Simulated trimmer position
display
(The trimmer moves to the
directly-pressed place.)
* Pressing the simulated trimmer
operation.
* The trimmer adjusting ranges are as follows: voltage output; about ±2.0V, and current
output; about ±1.6mA.
* The [Clear] button resets the Simulated trimmer position to the trimmer center point (0%).
: Output is produced in synchronization with the indicated value.
: Output is produced in synchronization with the sensor input value.
: Output is fixed at zero output (0V or 4mA).
: Output is fixed at full scale (10V or 20mA).
Option Setting (D/A OUTPUT)
Clear button
buttons continuously performs continuous
6 OPTION
Page 1
button, and then use the
button, so
Simulated trimmer
adjustment
UP ・ DOWN button
77
Chapter
6