Self Check; D/A Converter (Option) - Unipulse F377A Operation Manual

Digital indicator
Hide thumbs Also See for F377A:
Table of Contents

Advertisement

-STROBE Strobe pulses are output in synchronization with BCD data. Read data using the rising
      Example) Negative logic
BCDdata
OVER
0
STROBE
1

■Self check

The BCD input / output status can be checked from bit to bit.
Each input turns green when its ON state is recognized.
Each output turns on when the corresponding button is pressed.
How to set
SET.
6-2.

D/A Converter (Option)

A D/A converter is provided for obtaining analog output synchronized with the indicated value of
the F377A.
The analog output ranges are -10 - +10V output and 4 - 20mA constant-current output.
By using the D/A zero setting and D/A full scale setting functions, analog output can be obtained
between zero (0V, 4mA) and full scale (10V, 20mA) with respect to the predetermined digital value.
The output circuit and the main unit circuit are isolated. The resolution is 1/10000 to -10 - +10V
and 4 - 20mA, and the conversion rate is 2000 times/sec. The maximum voltage output range are
about ±11V, the maximum current output range are about 2.4 - 21.6 mA.
edges of the pulses. The BCD data output rate setting can be changed.
Strobe range
(duty50%)
System Setting
Option Setting (BCD OUT)
Voltage / current output terminals:
erminals to extract voltage or
current signals.
-10 to +10V and 4 to 20mA can
be obtained by the voltage output
and current output, respectively.
6 OPTION
Page 1