- Hysteresis operation
Indicated Value
HI Limit
LO Limit
+
0
-
HI Output
HI Status Display
LO Output
LO Status Display
Output ON
Status ON
OK Output
OK Status Display
■Near zero
By this function, it is detected that the indicated value is near zero.
Near-zero ON: | indicated value | ≦ near zero set value
Near-zero OFF: | indicated value |> near zero set value
Setting range: 00000 - 99999
Near-zero ON/OFF is closely related to the SI/F Print Out (stable) and Comparison timing.
For details, see the sections "SI/F Print Out" on page 40 and "Comparison timing" on page
43.
How to set
SET.
Work Setting
→
■Comparison timing
Set the operating condition of HI-LO limit comparison. Select the condition from the following.
ALL:
HI-LO limit comparison is performed continuously.
MD:
HI-LO limit comparison is performed when the indicated value is stable.
Set the stable parameters by the stable operation.
NZ:
HI-LO limit comparison is performed when the indicated value is not near zero.
Set the near zero parameters by the near zero operation.
MD+NZ: HI-LO limit comparison is performed when the indicated value is stable and not near zero.
HOLD:
HI-LO limit comparison is performed when the indicated value is held.
How to set
SET.
Work Setting
→
Output OFF
Output ON
Status OFF
Status ON
Output OFF
Status OFF
Status ON
Comp. Setting
→
Comp. Setting
→
5 SETTING OF FUNCTIONS
Hysteresis Range
Time
Output OFF
Status OFF
Output ON
Status ON
(Exsample :
HI Output and LO Output and OK output)
Status ON
Page 2
→
Page 2
→
Chapter
5
43