4 CALIBRATION
4-1.

Calibration Procedures

Follow the steps below to perform equivalent input calibration and actual load calibration.
Chapter
4
*
*
30
4-2.
Calibration Protect
4-3.
Calibration Selecttion
*
4-4.
Analog Input Selection
*
4-5.
Unit
*
4-6.
Zero Calibration
4-7.
4-8.
Equivalent
Actual Load
Calibration
Calibration
or
4-9.
Increment
4-10.
Digital Offset
4-11.
Digital Zero Limit
4-2.
Calibration Protect
...It is indispensable.
Release the calibration protection.
Set the calibration value No.
(Set "0" for use with only one selection.)(This step
may be omitted if there is no change.)
Set the analog input (Voltage・current) excitation
voltage according to the voltage・current output
device (sensor) used.
Set the unit of the values displayed.
Set the zero point of the voltage・current output
device (sensor) in no-load condition (with the
sensor unloaded).
Equivalent Input Calibration
Enter the output value and reading of the voltage・
current output device (sensor) .
At this time, also input the decimal point.
Actual Load Calibration
Enter the span (gain) point of the voltage・current
output device (sensor) by applying the load to the
sensor.
At this time, also input the decimal point.
Enter the minimum value of digital increments.
(This step may be omitted if there is no change.)
The calibrated value can be offset in advance.
(When not using, set "0".)
(This step may be omitted if there is no change.)
Set the load limit to allow digital zero.(This step
may be omitted if there is no change.)
Turn on the calibration protection for preventing
misoperation.