5 SETTING OF FUNCTIONS
5-13.
Measurement Work Selection
Select the method for specifying the measurement work.
How to set
SET.
5-14.
Control Input Selection
Select the method for specifying the control (SECTION signal, T/H signal, or GRAPH TRIG
signal) input.
Chapter
5
How to set
SET.
5-15.
B6 OFF Detection Wait
In hold controlled by the TH/signal alone, detection/hold section is assured during the set waiting
time. This is convenient for ignoring chattering parts.
Setting range: 0.00 - 1.00 sec.
Example) Sample hold
How to set
SET.
58
COM.:
Measurement work specification by external input becomes ineffective, and
only measurement work specification by communication (RS-232C, CC-Link
(option) or DeviceNet (option)) becomes effective.
EXT.:
Measurement work specification by communication (RS-232C, CC-Link
(option) or DeviceNet (option)) becomes ineffective, and only measurement
work specification by external input becomes effective.
System Setting
COM.: Control by external input becomes ineffective, and only control by communication
(RS-232C, CC-Link (option) or DeviceNet (option)) becomes effective.
EXT.:
Control by communication (RS-232C, CC-Link (option) or DeviceNet (option))
becomes ineffective, and only control by external input becomes effective.
System Setting
Hold point to be taken under normal
circumstances
Hold point displaced due
to chattering
Tracking section
Hold section
Chattering
OFF
(80msec)
T/H
ON
When the waiting time is 0.00 sec.
System Setting
Operation
Operation
t
Tracking section
Tracking section
OFF
T/H
ON
Operation
Page 3
Page 3
Hold section
Tracking section
0.10-sec.
assured section
Chattering operation is ignored
because the input is not recognized.
When the waiting time is 0.10 sec.
Page 3
t