5 SETTING OF FUNCTIONS
5-13.
Measurement Work Selection
Select the method for specifying the measurement work.
How to set
SET.
5-14.
Control Input Selection
Select the method for specifying the control (SECTION signal, T/H signal, or GRAPH TRIG
signal) input.
Chapter
5
How to set
SET.
5-15.
B6 OFF Detection Wait
In hold controlled by the TH/signal alone, detection/hold section is assured during the set waiting
time. This is convenient for ignoring chattering parts.
Setting range: 0.00 - 1.00 sec.
Example) Sample hold
How to set
SET.
58
COM.:
Measurement work specification by external input becomes ineffective, and
only measurement work specification by communication (RS-232C, CC-Link
(option) or DeviceNet (option)) becomes effective.
EXT.:
Measurement work specification by communication (RS-232C, CC-Link
(option) or DeviceNet (option)) becomes ineffective, and only measurement
work specification by external input becomes effective.
System Setting
→
COM.: Control by external input becomes ineffective, and only control by communication
(RS-232C, CC-Link (option) or DeviceNet (option)) becomes effective.
EXT.:
Control by communication (RS-232C, CC-Link (option) or DeviceNet (option))
becomes ineffective, and only control by external input becomes effective.
System Setting
→
+
Hold point to be taken under normal
circumstances
Hold point displaced due
to chattering
Tracking section
Hold section
Chattering
OFF
(80msec)
T/H
ON
When the waiting time is 0.00 sec.
System Setting
→
Operation
→
→
Operation
→
→
+
⇒
t
Tracking section
Tracking section
OFF
T/H
ON
Operation
→
→
Page 3
Page 3
Hold section
Tracking section
0.10-sec.
assured section
Chattering operation is ignored
because the input is not recognized.
When the waiting time is 0.10 sec.
Page 3
t