Self-Check - Unipulse F377A Operation Manual

Digital indicator
Hide thumbs Also See for F377A:
Table of Contents

Advertisement

8-3.

Self-Check

■Self-check
The self-check function includes a memory check to check the memory automatically for detecting
problems, a visual check to check the display visually, a touch panel key input check and an
external I/O check.
①LCD check
Check the display to see that it is free from color and display defects, etc. The screen changes from red,
green, blue, horizontal stripes, to vertical stripes. Touching the screen restores the previous screen.
How to set
SET.
②KEY check
Check the touch panel to see that it is free from unrecognized parts.
One touched square turns yellow. Pressing the
right of the screen restores the previous screen.
How to set
SET.
③MEM check
Check the memory to see that it is free from defects. The result is
displayed in a moment after the
screen is pressed. Or, check the version here.
How to set
SET.
④I/O check
Check the operations of the external input/output signals.
Each output turns on when the corresponding button is pressed.
Each input turns green when its ON state is recognized.
How to set
SET.
⑤DSP check
Check the backlight and status indicator lamp.
The backlight turns bright by pressing "
dark by pressing "
The backlight goes out when the "
after that, it recovers when the screen is touched.
The status indicator lamp lights in green and orange alternately.
How to set
SET.
Self Check
LCD Check
Self Check
KEY Check
at the lower right of the
Self Check
MEM Check
Self Check
I/O Check
" button.
" button is pressed, and
Self Check
DSP Check
8 SUPPLEMENTS
at the upper
" button, and turns
Chapter
8
87

Advertisement

Table of Contents
loading

Table of Contents