3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6
Note
Do not fill anything into the central opening (tumbler drain).
▶ Quit replenishing immediately if you hear an acoustic signal being issued
every second (10 sounds).
ð
Reservoir is full.
▶ Attach the tumbler holder and the tumbler filler.
Note
Check the O-ring on the tumbler filler for damage.
3.1.1 Replenish OXYGENAL 6 when using the water
bottle
Damage caused by defective water bottle.
Malfunctions or failures from a leaky water bottle.
▶ Do not wash the water bottle in a dish washer, and do not rinse it with wa-
ter hotter than 50 ℃.
▶ Do not use the water bottle if it is scratched, deformed or discoloured.
▶ Note the maximum useful life of the water bottle on the label.
Water bottle complete: (Mat. no. 0.710.4151)
Dosing attachment: (Mat. no. 1.002.0287)
Removing and cleaning the water bottle
▶ Slowly turn the water bottle counterclockwise to take it off.