KaVo Primus 1058 Life Care Instructions
Hide thumbs Also See for Primus 1058 Life:

Advertisement

Care instructions
Primus 1058 Life

Advertisement

Table of Contents
loading

Summary of Contents for KaVo Primus 1058 Life

  • Page 1 Care instructions Primus 1058 Life...
  • Page 2 Manufacturer: Distributed by: Kaltenbach & Voigt GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach www.kavo.com Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Target group..........................4 2 Setup methods according to DIN EN ISO 17664..................... 5 2.1 References............................... 5 2.2 General instructions for the reprocessing of KaVo treatment units............5 2.2.1 Reprocessing Overview Primus 1058 Life................... 8 2.3 Reprocessing of the surfaces of treatment unit / upholstery..............8 2.4 Preparation of the dentist unit........................
  • Page 4: User Instructions

    Care instructions Primus 1058 Life 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the treatment unit! 1.1.1 Abbreviations...
  • Page 5: Setup Methods According To Din En Iso 17664

    The preparation methods for the instruments and motors are found in the separate instructions for use that accompany the instrument and motor packages. 2.2 General instructions for the reprocessing of KaVo treatment units Please comply with the general reprocessing instructions at all times. Detailed prod‐...
  • Page 6 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Storage and transport KaVo recommends reprocessing as soon as possible after use. Preparations for cleaning ▶...
  • Page 7 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Automatic disinfection KaVo recommends washer disinfectors in accordance with EN ISO 15883-1, which are operated with alkaline cleaning agents with a pH value of max.
  • Page 8: Reprocessing Overview Primus 1058 Life

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery 2.2.1 Reprocessing Overview Primus 1058 Life Reprocessing items/medical devi‐ Cleaning Disinfection Sterilisation manual automatic manual...
  • Page 9: Preparation Of The Dentist Unit

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Cleaning and disinfecting the surfaces of treatment unit / upholstery CAUTION Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
  • Page 10 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Cleaning and disinfection of the handle TM table / cart ▶ Clean the handle with a soft cloth and water, and wipe-disinfect it with an ap‐...
  • Page 11: Reprocessing Of The Instrument Holder Of The Tm Table

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit Sterilisation of S-table handle and holder pad/anti-slip mat CAUTION Damage caused by sterilisation. ▶ Do not sterilise the white standard handle.
  • Page 12: Reprocessing Of The Instrument Arm S Table

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Pull the latches evenly outward and carefully remove the handpiece holder. ▶ Clean the handpiece holder under running water.
  • Page 13 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Fit the germ reduction attachment to the spittoon bowl. ▶ Take the handpieces off the hoses. ▶ Take the roller out of the instrument arm and then place the hose in the germ re‐...
  • Page 14: Setting Up The 3-Function And Multifunction Handpiece

    Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
  • Page 15 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Carefully wipe off contamination from the cannula tip with a soft Q-tip or soft cloth and water. CAUTION Damage caused by using a nozzle needle that is too long.
  • Page 16: Preparation Of The Instrument Hoses

    2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula.
  • Page 17: Servicing The Turbine Return Air Filter

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit 2.4.6 Servicing the turbine return air filter Cleaning the turbine return air filter Note The liquid collector must be checked weekly.
  • Page 18 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit ▶ Clean the surface with a soft cloth and mild liquid cleaning agent. Disinfection Exterior disinfection by hand WARNING Highly flammable gases-.
  • Page 19: Reprocessing Of The Assistant Unit

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant unit ▪ FD 322 (Dürr Dental) ▪ Incidin Liquid (Ecolab) Note The exterior of the camera and camera hose can be disinfected with chemical disin‐...
  • Page 20: Reprocessing Of Sieve Inserts And Suction Hoses

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit 2.5.3 Reprocessing of sieve inserts and suction hoses Note Check the sieve inserts daily and after each surgery. The sieve inserts must be re‐...
  • Page 21 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit Note Inspect the O-ring (Mat. no. 0.200.6128) for damage and exchange if necessary to prevent water or oxidation damage.
  • Page 22: Preparation Of The Spittoon

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit 2.6.3 Preparation of the spittoon CAUTION Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
  • Page 23: Manual Reprocessing Of The Water-Conducting System

    ▶ With regard to the "Water block, compact" assembly kit, please note that no disin‐ fection facility is installed in the unit, and take appropriate safeguards. KaVo rec‐ ommends to use the "Water block DVGW with integrated water disinfection facility in combination with KaVo OXYGENAL 6 (Mat.
  • Page 24: Rinsing Programme

    During normal operation of the treatment centre, the continuous germ reduction func‐ tion secures continuous germ reduction in the water conducting systems due to the automatic dosing of the disinfectant, KaVo OXYGENAL 6. The rinsing programme and the intensive germ reduction facilitate the automatic rins‐...
  • Page 25 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system ▶ In the case of swing-arms without a catching function, insert the swing-arm lock (Mat. no. 1.008.1511) to ensure that the handpiece is removed.
  • Page 26 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system ▶ For the rinsing programme and intensive germ reduction, press the seat adapter into one of the four rubber seats of the germ reduction attachment.
  • Page 27: Intensive Germ Reduction

    During intensive germ reduction, the affer‐ ent water ducts are automatically filled with a higher concentration of KaVo OXYGEN‐ AL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hy‐...
  • Page 28 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Note If the amalgam separator malfunctions during intensive germ reduction, the entire process is stopped. The bottom LED of the "HYDROclean" button flashes.
  • Page 29 Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Interrupt intensive germ reduction during the dwell time So that no one will have to remain with the device throughout the entire intensive germ reduction process, intensive germ reduction can be interrupted in program step 4 (e.g.
  • Page 30: Reprocessing And Automatic Hygiene Functions Of The Suction System

    The unit is now ready for use. Additional intensive germ reduction during period of non-use KaVo recommends carrying out an additional intensive germ reduction at universities during semester breaks. This flushes the parts of the unit and replaces the germ re‐...
  • Page 31: Hydroclean Function

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.10 Preparation of the amalgam separator 2.9.2 HYDROclean function The HYDROclean function is an automatic programme for hydro-mechanical cleaning of the drainage and separation systems within and outside the treatment unit. The cleaning process is carried out by the supply of fresh, clear water (limited rinsing).
  • Page 32: Reprocessing Kavo Screen One / Kavo Screen Hd

    Care instructions Primus 1058 Life 2 Setup methods according to DIN EN ISO 17664 | 2.11 Reprocessing KaVo Screen One / KaVo Screen HD Optimal care of the amalgam separation system includes the correct and regular use of the HYDROclean function and use of DEKASEPTOL gel.
  • Page 33: Replenish And Check The Disinfectant

    3 Replenish and check the disinfectant 3.1 Replenish OXYGENAL 6 The only permissible agent for use in KaVo water germ reduction systems is KaVo OXYGENAL 6 (Mat. no. 04893451). The OXYGENAL 6 consumption depends on the water consumption during the treatment of patients.
  • Page 34: Replenish Oxygenal 6 When Using The Water Bottle

    Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6 ▶ Quit replenishing immediately if you hear a signal sound being issued every sec‐ ond (10 sounds). ð Reservoir is full. ▶ Attach the tumbler holder and the tumbler filler.
  • Page 35 ▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing fa‐ cility. ▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXYGENAL 6 bottle. ▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
  • Page 36 In response to over-filling, the excess quantity above the tick mark is aspirated au‐ tomatically to ensure exact dosing. ▶ Place the water bottle in the KaVo OXYGENAL 6 bottle. Make sure that the spout of the KaVo OXYGENAL 6 dosing facility (nose) is situated right above the open‐...
  • Page 37: Check The Oxygenal 6 Concentration

    Care instructions Primus 1058 Life 3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after condensation of the water.

Table of Contents