Short instructions for use
Primus 1058 Life

Advertisement

Table of Contents
loading

Summary of Contents for KaVo Primus 1058 Life

  • Page 1 Short instructions for use Primus 1058 Life...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 (0) 7351 56-0 www.kavo.com Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Table of contents Table of contents Controls............................4 Dentist element TM/C table....................4 Dentist element S-table......................5 Assistant element........................5 Groups of keys........................6 Foot control..........................8 Moving the patient chair........................9 Positioning the dental chair manually..................9 Positioning the patient chair with the foot control..............9 Automatic positioning of patient chair..................
  • Page 4: Controls

    Short instructions for use 1 Controls 1 Controls CAUTION These brief instructions for use only contain the essential operating functions. It does not substitute for the Instructions for Use and Servicing that are part of the scope of delivery. ▶ The Instructions for Use and Servicing include all information and safety notes that must be observed to ensure intended use and safe operation of the product! ▶...
  • Page 5: Dentist Element S-Table

    Short instructions for use 1 Controls 1.2 Dentist element S-table Dentist element S-table Group of keys for menu selection Group of keys for the timer (MEMOspeed optional) Group of keys for the handpieces Group of keys for the dental chair Group of keys for illumination Group of keys for hygiene 1.3 Assistant element...
  • Page 6: Groups Of Keys

    Short instructions for use 1 Controls 1.4 Groups of keys Group of keys for the dental chair The keys of the assistant unit each have two functions and show two symbols. Assistant element key Dentist element key Name "Chair up" key "AP 0"...
  • Page 7 Short instructions for use 1 Controls Name Control element "Bowl rinsing" key Dentist element and assistant element "Intensive germ reduction" key Assistant element (optional) "HYDROclean" key Assistant element Group of keys for the handpieces/timer Name Control element "Preselected spray" button Dentist element "Direction of motor rotation"...
  • Page 8: Foot Control

    Short instructions for use 1 Controls 1.5 Foot control Item Name Function with handpiece Function with handpiece re‐ mounted moved ① "Spray pre-selection/AP" Moves the dental chair into Sets the spray pre-selection. footswitch automatic position. ② U-shaped switch Turns the safety shutoff On. Switches the footswitches to the "Chair motion"...
  • Page 9: Moving The Patient Chair

    Short instructions for use 2 Moving the patient chair 2 Moving the patient chair 2.1 Positioning the dental chair manually Feature Dentist element Assistant element The chair moves up. The chair moves down. The backrest moves upward. The backrest moves downward. 2.2 Positioning the patient chair with the foot control See also: 2 "Chair position"...
  • Page 10 Short instructions for use 2 Moving the patient chair Recalling an automatic position through the assistant element ▶ Briefly press the "AP" key. ð The LEDs of the "AP 0", " AP 1", " AP 2", "SP", and "LP" keys flash for approxi‐ mately four seconds.
  • Page 11: User Menu

    Short instructions for use 3 User menu 3 User menu The following options can be opened in the user menu: Option Feature Description Firmware Display current firmware version. Time of day Set time of day. Date Set date. Time display mode Setting the time of day display mode: •...
  • Page 12: Selecting And Setting An Option

    ▶ Press the "Save" key (S1) to exit from the user menu. ð The changed settings are saved. 3.2 Selecting and setting an option ▶ Press the "Decrease value" or the "Increase value" keys to select the desired pa‐ rameter or value. See also: 2 Instructions for use Primus 1058 Life...
  • Page 13: Switching Between Levels And Selecting A Dentist

    Short instructions for use 4 Switching between levels and selecting a dentist 4 Switching between levels and selecting a dentist Standby menu The unit starts in the standby menu. Available with MEMOspeed licence only: The device automatically switches to its standby menu when you close the handpiece menu and the patient communication menu.
  • Page 14 Short instructions for use 4 Switching between levels and selecting a dentist Note The device automatically saves the activation of level switching for the current den‐ tist. Note Level switching is deactivated using the same key combination as activation. ▶ Briefly press "Preselect level" key to select the level. Selection of dentist ▶...
  • Page 15: Setting The Timer

    Short instructions for use 5 Setting the timer 5 Setting the timer 5.1 Select the timer time ▶ Press the "Timer" key briefly to start or stop the timer. ð LED flashes while the timer counts down. The elapsed time is shown in the display on the top right. A beep is issued after the timer time is elapsed.
  • Page 16: Setting The Handpieces

    Short instructions for use 6 Setting the handpieces 6 Setting the handpieces 6.1 Adjusting functions Switch to the settings menu of the active element Level switching must be activated for the selection of a level. See also: 2 4 Switching between levels and selecting a dentist, Page 13 ▶...
  • Page 17 Short instructions for use 6 Setting the handpieces Setting the cooling level Requirement Settings menu for turbine is selected. ▶ Press the "Preselected spray" button. Feature No LED is on: No cooling One LED is on: Spray air cooling status Both LEDs are on: Spray cooling status Setting the direction of motor rotation Note...
  • Page 18: Turbine

    Short instructions for use 6 Setting the handpieces Define operating mode (PiezoLED only) Note The selection of the mode depends on the treatment method and the tip used. For information about the selection of an operating mode, please refer to the "Operat‐ ing modes P1 / P2 / P3 and E"...
  • Page 19: Multifunctional Handpiece

    Short instructions for use 6 Setting the handpieces The following settings can be changed: • Output intensity • Operating mode (P1 / P2 / P3 / E) ▪ Cooling status (no cooling / spray water cooling) Settings menu for minimal/maximal intensity 6.5 Multifunctional handpiece The following settings can be changed: ▪...
  • Page 20: Using Handpieces

    Short instructions for use 7 Using handpieces 7 Using handpieces ▶ Take the handpiece (e.g., turbine, motor, PiezoLED, etc.) from the holder. ð The handpiece is active. ▶ Press the foot pedal. ð The removed handpiece runs at the set speed or intensity. ▶...
  • Page 21: Operation Of The Operating Light Kavolux 540 Led (Optional)

    • Dimmed light: approx. 4,000 Kelvin; equivalent to the light of a halogen operating lamp ▪ Laser mode: Light mode, which has no negative influence on the KaVo KEY Laser III, the KEY Laser 3+ or the KaVo DIAGNOdent.
  • Page 22 Short instructions for use 8 Operation of the operating light KaVoLUX 540 LED (op‐ tional) Switch on COMPOsave mode ▶ Press the "Dim treatment light" key. ▶ Hold your hand in front of the sensor for 2 sec. ð The COMPOsave mode switches on. ð...
  • Page 23 Short instructions for use 8 Operation of the operating light KaVoLUX 540 LED (op‐ tional) Operation of the 3D joint ▶ Turn the switching ring to the left until it snaps into place. ð Treatment light can be rotated 45° to the left or 45° to the right. ▶...
  • Page 24: Using The Conexiocom (Optional)

    The function of the CONEXIOcom menu is to control the display of previously recorded and saved im‐ ages and videos. In order to use the function, the unit must have access to the data of the KaVo Soft‐ ware "CONEXIO" software. For details on the configuration, please refer to the "CONEXIO" installation instructions.
  • Page 25: Using The Patient Communication Menu With The Foot Control

    ▶ To activate a function, press the key below the info line briefly or long. Starting and exiting the patient communication menu ▶ Make sure that the treatment unit has access to the KaVo CONEXIO software. ð As soon as an imaging device is activated, the patient communication menu opens automati‐...
  • Page 26 Short instructions for use 9 Using the CONEXIOcom (optional) Item Feature Name ⑦ "LP" foot-operated button Select previous image/video ⑧ Foot pedal Save image/video: • Press briefly: Saves the image/video ▪ Press long: All images/videos are saved in the clipboard. Note If no patient is selected, the images stay in the "Swap Tray"...
  • Page 27: Hygiene

    Short instructions for use 10 Hygiene 10 Hygiene 10.1 Tumbler filling and bowl flushing The following settings can be changed: • Tumbler filling time ▪ Bowl rinsing time ▶ Press the "Tumbler filler" button briefly to start filling the tumbler. ▶...
  • Page 28 Short instructions for use 10 Hygiene S-table ▶ Lock the swing-arm into the front catching position to ensure that the handpiece is removed. ▶ In the case of swing-arms without a catching function, insert the swing-arm lock (Mat. no. 1.008.1511) to ensure that the handpiece is removed. ▶...
  • Page 29: Intensive Germ Reduction

    If an audible signal is issued every 10 seconds and the bottom LED of the "Intensive germ reduction" button flashes, the KaVo OXYGENAL 6 must be replenished. If a beep is issued during the filling process (10 times at a one-second interval), the con‐...
  • Page 30: Reprocessing The Suction System Including Suction Hoses

    Short instructions for use 10 Hygiene ▶ Perform the preparations for intensive germ reduction. See also: 2 10.2 Rinsing programme, Page 27 ▶ Press the "Intensive germ reduction" button. ð Intensive germ reduction starts and runs though several steps. The process can be followed on the display.
  • Page 31: Hydroclean Function

    Short instructions for use 10 Hygiene Application of DEKASEPTOL Gel ▶ Aspirate one tumbler of cold water with each suction hose. ▶ With the suction handpiece attached or by hand, press down the Dekaseptol dosing facility once each for the saliva ejector hose and the spray mist ejector. ▶...
  • Page 32: Replenishing Oxygenal

    Short instructions for use 10 Hygiene 10.6 Replenishing Oxygenal When the Oxygenal container is empty, a beep is issued every 10 seconds. ▶ Remove the tumbler filler ① and tumbler holder ②. ▶ Top off the Oxygenal (max. 1 litre). Note Do not fill through the centre opening (tumbler drain).
  • Page 33: Warning Messages

    Short instructions for use 11 Warning messages 11 Warning messages 11.1 Safety shut-off Safety shutoff for the standard patient chair Item No. Safety switch-off actuated LED on assistant element LED on dentist element ① Patient unit swung over the patient chair ②...
  • Page 34: Additional Warning Messages

    Short instructions for use 11 Warning messages 11.2 Additional warning messages Malfunction Cause Remedy A beep is issued every The Oxygenal container ▶ Refill the Oxygenal container. ten seconds and the "In‐ is empty. tensive germ reduction" LED (green) flashes. Ten beeps are issued.
  • Page 35: Service And Error Messages In The Stand-By Menu

    Short instructions for use 12 Service and error messages in the stand-by menu 12 Service and error messages in the stand-by menu If a status message is available, the standby menu shows an exclamation mark on selection key "S2" ①. ▶...
  • Page 36 Short instructions for use 12 Service and error messages in the stand-by menu Malfunction Cause Remedy Display shows: ID 69 Intensive germ reduc‐ ▶ Carry out an intensive germ reduction. tion must be done. See also: 2 Servicing instructions Display shows: ID 70 Dekaseptol empty.