KaVo Primus 1058 Life Care Instructions page 15

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Manual cleaning
Note
Combine manual cleaning with disinfection.
Cleaning the interior
▶ Leave the gripping sleeve and cannula on the instrument, and activate the
air and water for 60 seconds.
External cleaning
Danger of injury if the valve body is hot (heating elements, high-pres-
sure lamp).
Risk of burn injury.
▶ Switch off the unit at the main switch.
▶ Let the handpiece cool after long use.
Accessories required:
o
▪ Tap water 30
C ± 5
▪ Brush, e.g. medium-hard toothbrush
▶ Disassemble the gripping sleeve and cannula.
▶ Detach the key part from the gripping sleeve.
▶ Clean the key part, gripping sleeve and cannula by brushing them under
running water (at least tap water quality).
▶ Immediately remove contamination from materials used in the dental prac-
tice (impression materials, caustic chemicals).
▶ Carefully wipe off contamination from the cannula tip with a soft Q-tip or
soft cloth and water.
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
▶ Use a short nozzle needle (Mat. no. 1.004.4986) to clean the media exit
holes.
▶ Carefully remove contamination from the media exit ports using the short
nozzle cleaning needle.
CAUTION
o
C
NOTICE
15 / 44