2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
An additional intensive germ reduction must be carried out during extended
periods on non-use (> 4 weeks).
The need for intensive germ reduction is automatically displayed in the top LED
of the "intensive germ reduction" button at first start-up and after periods of
non-use (weekend, holidays, vacation, etc.).
If an audible signal is issued every 10 seconds and the bottom LED of the "In-
tensive germ reduction" button flashes, the KaVo OXYGENAL 6 must be replen-
ished. If a beep is issued during the filling process (10 times at a one-second
interval), the container is full.
See also:
Intensive germ reduction must be carried out:
Start intensive germ reduction
See also:
Step
7.0
6.0
5.0
2 Replenish OXYGENAL 6
Note
The suction device cannot be used during the HYDROclean step (4.6).
Note
If the amalgam separator malfunctions during intensive germ reduction, the
entire process is stopped. The bottom LED of the "HYDROclean" button
flashes.
▪ At first start-up
▪ Before, during and after prolonged downtimes (more than 4 weeks)
▪ Over the weekend
▪ During vacations
▪ After treatment of patients at high risk
▶ Press the "Intensive germ reduction" button for more than 4 seconds until
you hear an acoustic signal and the upper LED of the "intensive germ re-
duction" button flashes. The display switches to the intensive germ reduc-
tion menu.
▶ Perform the preparations for intensive germ reduction.
2 2.8.1.1 Preparation for rinsing program and intensive germ reduction, Page
25
▶ Press the "Intensive germ reduction" button.
ð
This starts up the intensive germ reduction, which proceeds through several
stages. The process can be followed on the display. The major steps are lis-
ted below.
Description
Germ reduction of the triple-function and multifunctional handpiece
and the spittoon bowl
Duration: 60 seconds
Germ reduction of the handpieces
Duration: 30 seconds each
Germ reduction of the tumbler
Duration: 40 seconds
29 / 44
Care instructions Primus 1058 Life