Removing The Sonicflex Tip; Regulating The Spray; Check The Amount Of Water - KaVo SONICflex 2000 N Instructions For Use Manual

Table of Contents

Advertisement

Instructions for use SONICflex 2000 N - 0.571.5020
5 Operation | 5.4 Removing the SONICflex tip
Note
When the SONICflex is in the holder, the torque wrench should be placed on the tip
to protect against injury.

5.4 Removing the SONICflex tip

▶ Insert/attach the torque wrench in/on the SONICflex and unscrew the tip anti-
clockwise.

5.5 Regulating the spray

▶ Rotate the spray ring on the MULTIflex (LUX) / MULTIflex LED coupling in order
to regulate the fraction of water. The water volume can be regulated by selecting
different stop positions.
ð Turn clockwise to reduce the water volume.
ð Turn counterclockwise to increase the water volume.

5.6 Check the amount of water

Overheating of the tooth due to lack of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 30 cm
CAUTION
14 / 26
/min!